It is unlawful to practice medical physics in Florida without a license (diagnostic, therapeutic, nuclear or health physics). Licenses renew every two years with continuing education.
Source: Florida Statutes 483.901; FAC 64B23, checked 2026-11-26
Facilities register x-ray machines with the Department of Health, which inspects them on a schedule from yearly (mammography, therapy) to every five years (dental, podiatry).
Source: Florida DOH Bureau of Radiation Control, checked 2026-11-26
Anyone in the business of providing medical physics services in Hawaii must hold a license from the Department of Health and have at least one qualified medical physicist.
Source: Hawaii Administrative Rules ch. 11-45, checked 2026-11-26
Before registering a dental facility, the owner has a shielding design performed by a qualified expert and submits it to the state with the registration.
Source: Maine CDC Radiation Control Program, dental facility requirements, checked 2026-11-26
A radiation shielding plan review is submitted for new x-ray rooms, except for machines limited to intraoral dental, panoramic, bone densitometry and a few non-medical uses.
Source: MIOSHA Radiation Safety Section FAQ, checked 2026-11-26
Practicing medical physics or using the title 'licensed medical physicist' in New York requires a license from the State Education Department.
Source: NY Education Law article 166 (section 8705), checked 2026-11-26
People practicing medical physics in Texas need a license (or temporary license) from the Texas Medical Board: CAMPEP degree, specialty board certification and the jurisprudence exam.
Source: Texas Occupations Code ch. 602; 22 TAC ch. 601, checked 2026-11-26
A new or relocated mammography unit, or one that had a major repair, must pass a mammography equipment evaluation by a qualified medical physicist before it is used on patients.
Source: 21 CFR 900.12(e)(10), checked 2026-11-26
Suppliers of the technical component of CT, MRI, nuclear medicine and PET paid under the Medicare physician fee schedule must be accredited by a CMS-approved organization (ACR, IAC or The Joint Commission). Accreditors require medical physicist performance evaluations.
Source: 42 CFR 414.68; MIPPA section 135(a), checked 2026-11-26
Every FDA-certified mammography facility needs an on-site survey by a qualified medical physicist at least once a year (FDA treats 10 to 14 months after the previous survey as on time). The physicist sends a signed, dated report within 30 days; FDA inspectors review it at the annual MQSA inspection.
Source: 21 CFR 900.12(e)(9); FDA MQSA policy guidance, checked 2026-11-26
A qualified medical physicist must evaluate the performance of each CT unit at least annually; ACR and CMS site surveys check the reports.
Source: ACR CT Accreditation Program requirements; ACR validation site survey toolkit, checked 2026-11-26
ACR accreditation requires an annual performance survey by a qualified medical physicist (or MR scientist for MRI), with documented physicist qualifications.
Source: ACR personnel requirements for medical physicists and MR scientists, checked 2026-11-26
Each CT unit is evaluated at least annually (image quality items, and measured CTDIvol within 20% of the console value); adult and pediatric brain and abdomen dose protocols are reviewed yearly; physicist qualifications are documented. Dental cone-beam CT is excluded.
Source: Joint Commission diagnostic imaging standards (EC.02.04.03, HR.01.02.01), as reported in 2015-2016, checked 2026-11-26
A medical physicist (or MRI scientist for MRI) evaluates each MRI, nuclear medicine and PET system at least annually, and each image acquisition monitor.
Source: Joint Commission diagnostic imaging standards, as reported in 2015-2016, checked 2026-11-26
The physicist must meet MQSA's initial and continuing education and experience requirements for a qualified medical physicist before working independently.
Source: 21 CFR 900.12(a)(3), checked 2026-11-26
New 3D CBCT or cephalometric installations, and changes to their rooms, need a scaled drawing and shielding evaluation by a qualified expert.
Source: Vermont Department of Health dental x-ray inspection guide, checked 2026-11-26
A shielding plan prepared by a qualified expert is submitted to the Department of Health for review before installation (standard fee $977; a higher fee applies if equipment is installed before review). Dental, veterinary and podiatry facilities do not submit shielding calculations, but must still meet the radiation safety standards.
Source: WAC 246-225A; Washington DOH x-ray facility plan review, checked 2026-11-26