Anyone in the business of providing medical physics services in Hawaii must hold a license from the Department of Health and have at least one qualified medical physicist.
Source: Hawaii Administrative Rules ch. 11-45, checked 2026-11-26
A new or relocated mammography unit, or one that had a major repair, must pass a mammography equipment evaluation by a qualified medical physicist before it is used on patients.
Source: 21 CFR 900.12(e)(10), checked 2026-11-26
Suppliers of the technical component of CT, MRI, nuclear medicine and PET paid under the Medicare physician fee schedule must be accredited by a CMS-approved organization (ACR, IAC or The Joint Commission). Accreditors require medical physicist performance evaluations.
Source: 42 CFR 414.68; MIPPA section 135(a), checked 2026-11-26
Every FDA-certified mammography facility needs an on-site survey by a qualified medical physicist at least once a year (FDA treats 10 to 14 months after the previous survey as on time). The physicist sends a signed, dated report within 30 days; FDA inspectors review it at the annual MQSA inspection.
Source: 21 CFR 900.12(e)(9); FDA MQSA policy guidance, checked 2026-11-26
A qualified medical physicist must evaluate the performance of each CT unit at least annually; ACR and CMS site surveys check the reports.
Source: ACR CT Accreditation Program requirements; ACR validation site survey toolkit, checked 2026-11-26
ACR accreditation requires an annual performance survey by a qualified medical physicist (or MR scientist for MRI), with documented physicist qualifications.
Source: ACR personnel requirements for medical physicists and MR scientists, checked 2026-11-26
Each CT unit is evaluated at least annually (image quality items, and measured CTDIvol within 20% of the console value); adult and pediatric brain and abdomen dose protocols are reviewed yearly; physicist qualifications are documented. Dental cone-beam CT is excluded.
Source: Joint Commission diagnostic imaging standards (EC.02.04.03, HR.01.02.01), as reported in 2015-2016, checked 2026-11-26
A medical physicist (or MRI scientist for MRI) evaluates each MRI, nuclear medicine and PET system at least annually, and each image acquisition monitor.
Source: Joint Commission diagnostic imaging standards, as reported in 2015-2016, checked 2026-11-26
The physicist must meet MQSA's initial and continuing education and experience requirements for a qualified medical physicist before working independently.
Source: 21 CFR 900.12(a)(3), checked 2026-11-26